Search results for "Data acquisition board"
showing 4 items of 4 documents
A PC-based wattmeter for accurate measurements in sinusoidal and distorted conditions: Setup and experimental characterization
2012
This paper presents the development of a high-accuracy sampling wattmeter, which can be employed with both sinusoidal and distorted input signals. The proposed personal computer (PC)-based system is a simple, portable, and low-cost solution because it is based on two commercial data acquisition boards connected to a common PC. Owing to this solution, the voltage is sensed directly, avoiding expensive voltage dividers, while the current is sensed by means of a noninductive current shunt. Commercial software is also used for data processing and implementation of the virtual instrument. The setup of the PC-based wattmeter is described, with a detailed description of the correction strategy of …
Development of a high-accuracy PC-based wattmeter with commercial data acquisition boards
2011
This paper presents the development of a high accuracy sampling wattmeter, which makes use of two commercial data acquisition boards (DAQs) connected to a common personal computer. Commercial software is also used for data processing and implementation of the virtual instrument. The set up of the PC-based wattmeter is described and a preliminary evaluation of its metrological performances is made by means of two series of experimental tests, which were carried out with an electric power standard and a high performance multifunction calibrator. © 2011 IEEE.
A PC-based wattmeter for high accuracy power measurements
2010
This paper proposes the use of a commercial data acquisition board (DAQ) connected to a common personal computer for high accuracy power measurements. In the paper the set up of the PC-based wattmeter is described and an experimental evaluation of its metrological performances is made by means of the comparison with a simultaneous coherent digital sampling wattmeter realized with two high-precision DMMs Hewlett Packard Model 3458 A. In both cases the power source is a Fluke 6100A/E electrical power standard. © 2010 IEEE.
Uncertainty evaluation in power measurements with commercial data acquisition boards
2011
The paper deals with the metrological characterization of a digital wattmeter developed by using a personal computer, a shunt, two commercial data acquisition boards and commercial software. The uncertainty of the active power measurements is assessed individually characterizing the voltage channel, the current channel and the phase displacement measurement between the two channels. In particular, the uncertainty on the voltage and current rms measurements is evaluated by using an already proposed approach based on the Monte Carlo method and on the usage of five parameters, while the uncertainty on the phase angle is experimentally estimated. In order to validate the proposed approach for t…